Skip to main content

Table 2 Comparison of stability between instruments of three methods: A. internal self mean ratio; B. ratios based on an external assay; and C raw instrument data. Internal self mean (A) is the most reliable. Using external ratios, (2C) is a close second, and raw readings, (2C) show the greatest deviation between instruments.

From: Variance in multiplex suspension array assays: intraplex method improves reliability

 

Ratios N = 32 for all.

Mean z score

Median z score

Max z

Min z

A

Ratios on internal self mean

    
 

Rh. Sera 1:50

1.07

0.84

2.08

0.52

 

Rh. Sera 1:100

0.96

0.92

1.21

0.76

 

Rh. Sera 1:200

1.86

2.31

3.50

0.14

B

Ratios on real external mean (SRV/SFV mean and SFV/SRV mean)

    
 

Rh. Sera 1:50

1.70

1.82

2.50

0.50

 

Rh. Sera 1:100

1.44

1.44

3.83

0.27

 

Rh. Sera 1:200

16.38

11.97

49.47

4.59

C

Raw inter instrument comparisons

    
 

Rh. Sera 1:50

1.19

1.09

1.75

0.61

 

Rh. Sera 1:100

8.14

7.91

9.69

6.98

 

Rh. Sera 1:200

46.04

43.89

65.40

26.52